duoPUR Acid Purification System: Technical Details

Easy Operation

  1. Fill the acid/water container with standard-grade reagent or water.
  2. Program the distillation cycle.
  3. After preset time, unit switches off.
  4. If acid/water are completely distilled before the end of the cycle, a thermal switch will automatically turn the unit off.
duoPUR schematic
Schematic of single acid still

Technical Information

subPURduoPUR
Number of distillation units12
Heating elements2 with gold diffuser4 with gold diffuser
Power rate1250W1250W
Power controlMicroprocessor in 1% increments from 0 to 100%
Temperature controlThermal switch to prevent overheating in absence of acids
Water coolingAutomatic ON/OFF through solenoid valve
Control terminalLCD display, 2x20 lines with up to 10 programs in memory
Dimensions55 x 35 x 60 cm
(21 x 14 x 24 in.)
Weight20 kg (44 lbs.)22 kg (48 lbs.)

Comparison of trace metal contamination in select high-purity nitric acids

Note: For additional information, see our Acid Purity technical note, with comparative data provided by actual DuoPUR customers.

duoPUR
double distilled
Fisher
OptimaTM
Baker
Ultrex® II
Ti5.5 ± 1.1< 20< 100
V0.8 ± 0.4< 1< 20
Cr3.1 ± 0.7< 10< 50
Mn1.4 ± 0.4< 1< 20
Co0.5 ± 0.3< 1< 20
Ni8.1 ± 3.0< 10< 100
Cu10 ± 2.5< 2< 50
Zn12 ± 3.2< 2< 100
Concentration shown in pg/g.

Trace Element Concentration at 300W of the Most Common Digestion Reagents

HNO3HClH2O
Al< 0.05< 0.08< 0.05
Cd0.00080.010.01
Cu0.03< 0.080.04
Fe0.10.50.3
Pb0.030.050.02
Concentration shown in ng/mL (ppb).

Typical Yields

PowerHNO3HClH2OH2SO4
200W51.5346.9566.0-
400W145.71156.52174.0-
600W235.71260.85270.013.0
800W325.71375.65360.026.09
1000W420.00443.48456.039.13
Yields shown in mL/h.